Description and Specification for Lab Tenders
Flamephoto Meters.
Simple and easy to operate, with narrow-band interference filters and a dual-channel display reading Na and K at a single aspiration.
Flame photometer unit:
· Filters: Na and K (Ca and Li optional)
· Range: Na 0–100 ppm, K 0–100 ppm, Ca 20–100 ppm, Li 10–100 ppm (also available in meq/l: Na 0–200, K 0–100, Ca 0–2, Li 0–10)
· Sensitivity: Na 5 ppm, K 5 ppm, Ca 10 ppm, Li 10 ppm
· Display: 2½-digit bright LED (dual on the dual-channel model)
· Accuracy: ±2% up to 40 ppm, ±5% above 40 ppm
· Repeatability: ±2 counts
· Detector: silicon photodiode
· Filters: narrow-band interference glass filters
· Nebuliser: black bakelite, axial-flow type
· Flame system: LPG and dry, oil-free air
· Warm-up time: 10 minutes
· Power: 230 V ±10%, AC, 50 Hz
· Size (L x B x H): 355 x 250 x 210 mm (single) / 455 x 250 x 210 mm (dual)
· Weight: approx. 6.5 kg (single) / 7.5 kg (dual)
Compressor unit: oil-free mini compressor with pressure regulation; LPG combustion controlled by a precision regulator; power 230 V ±10%, AC, 50 Hz; size 275 x 192 x 240 mm; weight approx. 9 kg.
Accessories: instrument with Na and K filters, instruction manual, dust cover, chimney and air tubing.
We are manufacturers and suppliers of Flame Photometers, serving Algeria, Argentina, Australia, Bahrain, Bangladesh, Belgium, Bhutan, Brazil, Cambodia, Cuba, Cyprus, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, Georgia, Germany, Ghana, Hungary, Iceland, Italy, Japan, Jordan, Kenya, Kuwait, Malaysia, the Philippines, Tanzania and Uganda.